| u | Operating Modes | VVI, VOO (over drive :200to350 ppm) |
| u | R-Wave Sensitivity | Variable from 1mV to 20mV (2ms/13 ms triangular test pulse) scale accuracy : 1 to 10 mV ±10% 10 to 20 mV ±20% |
| u | Input Impedance | 8k ohm ± 10% |
| u | Defibrillation Protection | Suppression Diode |
| u | R-Wave Filter HF-Filter | High quality filter for acceptance of R-Wave specified 22Hz frequency components for suppression of high frequency Interference pulses |
| u | Output Pulse | Polarity: cathodic, capacitivity coupled form: biphasic asymmetric width :1.5 ms ± 0.03ms |
| u | Output Pulse Amplitude | stabilized output variables from 5V to 15V scale accuracy ±5% |
| u | Detection of Output Pulse | Red LED flashes synchronously to every stimulation pulse |
| u | Stimulation Frequency | variable from 30to 180ppm scale accuracy: 30-100ppm±5% >100ppm ±10% |
| u | Run Away Protection | 200ppm±15ppm |
| u | Refractory Period | (0.375*Pulse interval )ms at nominal setting :313ms±5% |
| u | Interference Recognition | Input pulse sequence <200ms(>4.5 Hz) cause the system to switch over to the asynchronous stimulation frequency |
| u | Battery Control | if green LED is on battery should be changed if green LED is off Battery is good |
| u | Battery | 9 Volt 1 no. |
| u | Dimension | 60mm*115mm*20mm, plastic housing |
| u | Weight | 185g(incl. battery) |
| u | Connection of Leads | Security sockets;1.8 to 2.3 mm diameter |
| u | Delivery Volume | SSB 100 Incl. battery |