Pacemaker
Pinnacle 8820 SSI (Multiprogrammable)

Pinnacle pacemaker is 8 Parameter multi-programmable Implantable Pulse Generator. Pinnacle has Digital Algorithm for automatic threshold measurement facility for tension free pacing even if the threshold is changing between acute and chronic levels. Once automatic feature is enabled the pacemaker itself settles down to an optimum pacing amplitude. When a loss of capture is detected pacemaker immediately increases its amplitude up to 4.5 Volts to maintain pacing, after 4.5Volts Automatic feature is disabled and pacing polarity is changed to Unipolar for effective pacing (eliminating defect caused by outer lead fracture).
Specifications
Programmable Parameter Values
u Pacing Modes VVI, VVT, VOO, OVO, AAI, AAT, AOO, OAO
u Auto Measure Switch ON / OFF
Automatic Ventricular Capture
u Basic Pacing Rates From 30 to 120 ppm in steps of 2 ppm
u Pulse Widths 21 values from 0.06 ms to 1.50 ms
u Pulse Amplitudes 32 values from 0.2 to 7.5 V
u Refractory Periods From 203 to 438 ms in steps of 15.625 ms
u Atrial Sensitivities From 0.5 to 7.5 mV
Ventricular Sensitivities From 0.5 to 7.5 mV
u Hysteresis From 0 to 20% in steps of 5%
u Pacing Polarities Unipolar / Bipolar
u Sensing Polarities Unipolar / Bipolar
u Upper rate in trigger mode From 80 to 183 ppm
Capture Management Algorithm
Provision for rate hysterisis, night low rate
Tachycardia Treatment
u Temporary programming Asynchronous pacing with rates from 30 to 427 ppm
Magnetic Response
u Mode VOO
u Rate BOL (begin of life): 95 ppm
ERI (elective replacement): 85 ppm
End of Service Indicators
ERI (Elective replacement): When the remaining capacity of the battery is between 4 and 8%, the pacing rate switches to 85 ppm when a magnet is applied.

EOL (End of Life): When the remaining capacity of the battery is under 4% of the initial capacity, the pacing rate slows 10 ppm.
Rate Drop – ON / OFF
u 1st detection 12 beats (1 to 30 beats)
u 2nd detection 3 beats (1 to 7 beats)
u RD Rate 96 ppm (80 to 146 ppm)
u Duration 180 beats (30, 60, 90, 120, 150, 180, 210, 240 beats)
u Upper limit 64 ppm (50 to 100 ppm)
u Lower limit 50 ppm (40 to 80 ppm)
Power Source
Battery Chemistry: Lithium – Iodine WG 9085
Initial Voltage: 2.8 V.
Maximum Available Capacity: 0.98 A-hr.
Physical Characteristics
Dimensions: 49 x 38 x 7 mm
Mass: 24 g.
Case Material: Titanium
Connectors: 3.2mm, IS-1
Service Life
The expected life of the generator, pacing 100% at 70 ppm, with 2.5 V amplitude and pulse width of 0.4 ms, is 10.2 years.
Telemetry Data
Real time telemetry with holter analysis trends.
Parameters: All the programmable parameters.
Identification: Model and serial number
Battery condition: Voltage, ERI and EOL indicators
Threshold measurements
Statistics: Keeps track of pacing and sensing events, as well as ventricular premature events. Frequency Histogram.
Lead Impedance
Marker Pulses.
Temporary Programming
Protections
Noise detection: When the pacemaker senses events with frequency exceeding 11 Hz, it switches to VOO mode.
Run Away Limit: The pacemaker includes an antirunaway circuit to avoid the generation of stimuli with a rate over 200 ppm (except in temporary programming).
Programming
The pacemaker is programmed using a microcomputer (usually a Laptop) that runs Windows operating system. It is required to use ALUS programming interface and Wand for programming custom made for Pacetronix Pacemakers. It also permits the measurement of the stimulation and sensing thresholds as well as the generation of printed reports.