Programmable Parameter Values |
|
| u | Pacing Modes | VVI, VVT, VOO, OVO, AAI, AAT, AOO, OAO | | u | Auto Measure Switch | ON / OFF Automatic Ventricular Capture | | u | Basic Pacing Rates | From 30 to 120 ppm in steps of 2 ppm | | u | Pulse Widths | 21 values from 0.06 ms to 1.50 ms | | u | Pulse Amplitudes | 32 values from 0.2 to 7.5 V | | u | Refractory Periods | From 203 to 438 ms in steps of 15.625 ms | | u | Atrial Sensitivities | From 0.5 to 7.5 mV
| | Ventricular Sensitivities | From 0.5 to 7.5 mV | | u | Hysteresis | From 0 to 20% in steps of 5% | | u | Pacing Polarities | Unipolar / Bipolar | | u | Sensing Polarities | Unipolar / Bipolar | | u | Upper rate in trigger mode | From 80 to 183 ppm | | Capture Management Algorithm | | | Provision for rate hysterisis, night low rate | | |
|
Tachycardia Treatment |
|
| u | Temporary programming | Asynchronous pacing with rates from 30 to 427 ppm | |
|
Magnetic Response |
|
| u | Mode | VOO | | u | Rate | BOL (begin of life): 95 ppm ERI (elective replacement): 85 ppm | |
|
| End of Service Indicators |
|
ERI (Elective replacement): When the remaining capacity of the battery is between 4 and 8%, the pacing rate switches to 85 ppm when a magnet is applied. EOL (End of Life): When the remaining capacity of the battery is under 4% of the initial capacity, the pacing rate slows 10 ppm. |
Rate Drop – ON / OFF | | | u | 1st detection | 12 beats (1 to 30 beats) | | u | 2nd detection | 3 beats (1 to 7 beats) | | u | RD Rate | 96 ppm (80 to 146 ppm) | | u | Duration | 180 beats (30, 60, 90, 120, 150, 180, 210, 240 beats) | | u | Upper limit | 64 ppm (50 to 100 ppm) | | u | Lower limit | 50 ppm (40 to 80 ppm) | | |
|
| Power Source |
|
Battery Chemistry: Lithium – Iodine WG 9085 Initial Voltage: 2.8 V. Maximum Available Capacity: 0.98 A-hr. |
|
| Physical Characteristics |
|
Dimensions: 49 x 38 x 7 mm Mass: 24 g. Case Material: Titanium Connectors: 3.2mm, IS-1 |
|
| Service Life |
|
| The expected life of the generator, pacing 100% at 70 ppm, with 2.5 V amplitude and pulse width of 0.4 ms, is 10.2 years. |
|
|
| Telemetry Data |
|
Real time telemetry with holter analysis trends. Parameters: All the programmable parameters. Identification: Model and serial number Battery condition: Voltage, ERI and EOL indicators Threshold measurements Statistics: Keeps track of pacing and sensing events, as well as ventricular premature events. Frequency Histogram. Lead Impedance Marker Pulses. Temporary Programming |
|
| Protections |
|
Noise detection: When the pacemaker senses events with frequency exceeding 11 Hz, it switches to VOO mode. Run Away Limit: The pacemaker includes an antirunaway circuit to avoid the generation of stimuli with a rate over 200 ppm (except in temporary programming). |
|
| Programming |
|
| The pacemaker is programmed using a microcomputer (usually a Laptop) that runs Windows operating system. It is required to use ALUS programming interface and Wand for programming custom made for Pacetronix Pacemakers. It also permits the measurement of the stimulation and sensing thresholds as well as the generation of printed reports. |
|