Pacemaker
Pinnacle-R SSIR + Model 297
Specifications
Programmable Parameters  Values
u Pacing Modes
VVI, VVT, VOO, OVO, AAI, AAT, AOO, OAO, VVIR, VVTR, VOOR, AAIR, AATR, AOOR
u Basic Pacing Rates From 32 to 120 ppm in steps of 2 ppm
u Pulse Widths 21 values from 0.07 ms to 1.50 ms in steps of 0.03 ms
u Pulse Amplitudes 36 values from 0.2 to 7.5 V
u Refractory Periods From 200 to 300 ms in steps of 15.625 ms
u Sensitivities From 0.5 to 11.0 mV
u Hysterisis
2 ppm to 40 ppm in steps of 2 ppm
u Pacing Polarities
Unipolar / Bipolar
u Sensing Polarities
Unipolar / Bipolar
u Upper rate in trigger mode
From 80 to 180 ppm in steps of 2 ppm
 
 Rate Response
u Upper Rate From 80 to 180 ppm in steps of 2 ppm
u Response to Activity From 1 to 16
u Reaction Times (Up Time) From 10 to 60 s, in steps of 10 s, to increment the rate 80 ppm
u Recovery Times (Down Time) From 1 to 10 m, in steps of 1 m, to decrement the rate 80 ppm
u Automatic Refractory Change When programmed, if the rate is greater than 120 ppm, the refractory period switches to 250 ms.

Tachycardia Treatment
u Temporary programming Asynchronous pacing with rates from 32 to 380 ppm

 Magnetic Response
u Mode VOO
u Rate BOL (begin of life): 95 ppm
ERI (elective replacement): 85 ppm

End of Service Indicators

ERI (Elective replacement): When the remaining capacity of the battery is between 4% and 8%, the pacing rate switches to 85 ppm when a magnet is applied.

EOL (End of Life): When the remaining capacity of the battery is under 4% of the initial capacity, the pacing rate slows 10 ppm.


Power Source
Battery Chemistry Lithium – Iodine WG 8711
Initial Voltage 2.8 V.
Maximum Available Capacity 0.82 A-hr.

Physical Characteristics
Dimensions 49 x 38 x 7 mm
Mass 24 g.
Case Material Titanium
Connectors 3.2mm, IS-1/VS-1

Service Life

The expected life of the generator, pacing 100% at 70 ppm, with 2.5 V amplitude and pulse width of 0.488 ms, is 10.2 years.


Telemetry Data
Parameters All the programmable parameters
Identification Model and serial number
Battery condition Voltage, ERI and EOL indicators
Statistics Keeps track of pacing and sensing events, as well as ventricular premature events. Frequency Histograms
Lead Impedance
Marker Pulses
Threshold Test
Rate Responsive Wizard

Protections

Noise detection: When the pacemaker senses events with frequency exceeding 11 Hz, it switches to VOO mode.

Run Away Limit: The pacemaker includes an anti-runaway circuit to avoid the generation of stimuli with a rate over 200 ppm (except in temporary programming).


Programming

The pacemaker is programmed using a microcomputer (usually a Notebook) that runs Windows operating system. It also permits the measurement of the stimulation and sensing thresholds as well as the generation of printed reports.