Pacemaker
Tiranga 302 BI/ 301 Uni 3 Parameter Programmable
Pacemaker
Specifications
Programmable Parameters  Values
u Pacing Modes
VVI, VVT, VOO, OVO, AAI, AAT, AOO, OAO
u Basic Pacing Rates From 30 to 120 Min-1 in steps of 2 Min-1
u Pulse Widths 0.40 ms
u Pulse Amplitudes 2.5V, 3.6V, 5.0V
u Refractory Periods 203, 250, 328, 406 ms Automatic
u Sensitivities 0.5 to 4.0 in steps of 0.5 mV
u Hysterisis
10% Factory Set
u Pacing Polarities
Unipolar for Model 301 & Bipolar for Model 302
u Sensing Polarities
Unipolar for Model 301 & Bipolar for Model 302
u Upper rate in trigger mode
120 Min-1
 
 Magnetic Response
u Mode VOO
u Rate BOL (begin of life): 95 ppm
ERI (elective replacement): 85 ppm

End of Service Indicators

ERI (Elective replacement): When the remaining capacity of the battery is between 4% and 8%, the pacing rate switches to 85 ppm when a magnet is applied.

EOL (End of Life): When the remaining capacity of the battery is under 4% of the initial capacity, the pacing rate slows 10 ppm.

ERI and EOL also Displayed when Interrogated by PCP+ Programmer.


 Power Source
u Battery Chemistry Lithium – Iodine Li 3150
u Initial Voltage 2.8 V.
u Maximum Available Capacity 1.28 A-hr.

 Physical Characteristics
u Dimensions 44 x 38 x 7.4 mm
u Mass .26 g
u Case Material Titanium
u Connectors 3.2mm, IS-1

Service Life

The expected life of the generator, pacing 100% at 70 ppm, with 3.6 V amplitude and pulse width of 0.4 ms, is 9.2 years.

The expected life of the generator, pacing 100% at 60 ppm, with 2.5 V amplitude and pulse width of 0.4 ms, is 12.6 years.


Telemetry Data

Parameters: All the programmable parameters.

Identification: Serial number

Battery condition: Voltage, ERI and EOL indicators Threshold test by Pulse width and sensing available.

Statistics: Keeps track of pacing and sensing events, as well as ventricular premature events.

Marker Pulses.

Protections

Noise detection: When the pacemaker senses events with frequency exceeding 11 Hz, it switches to VOO mode.

Run Away Limit: The pacemaker includes an antirunaway circuit to avoid the generation of stimuli with a rate over 200 ppm (except in temporary programming).

Programming

The pacemaker is programmed using a Laptop that runs Windows 98 / 2000 / XP operating system. The programming system includes Log of Pacemaker History. It also permits the measurement of the stimulation thresholds using Pulse width. Top